Where each pressure is measured, what an alarm usually means, and what to check first.
Four pressures are measured on most machines: access (P1, before the blood pump, negative), pre-filter (P2), effluent (P3) and return (P4). Pick an alarm below to highlight where it comes from.
Which alarm?
Choose an alarm to see likely causes and what to check.
Access pressure too negative
The blood pump is pulling harder than the catheter can supply.
Likely causes
Catheter arterial lumen against the vessel wall, kinked, or malpositioned
Patient position (hip flexion for femoral lines, coughing, agitation)
Clot or fibrin sheath at the catheter tip
Hypovolemia or blood flow set too high for the catheter
Clamp left closed on the access line
What to check
Check clamps and kinks, reposition the patient
Flush the arterial lumen
Lower blood flow temporarily
Reversing the lines can help short-term but increases recirculation
Repeated alarms: assess catheter position and consider exchange
Access pressure less negative than usual / near zero
Less suction than expected at the same blood flow.
If disconnection is possible, stop the blood pump and check before restarting.
Likely causes
Access line disconnected or separated from the catheter or pressure pod
Blood pump stopped or running slower than set
Improved catheter flow after repositioning (benign)
What to check
Check the access connection first: an open line can draw in air
Confirm the pump is running at the set rate
Return pressure too high
Resistance to returning blood to the patient.
Likely causes
Kinked or clamped return line
Clot in the return (deaeration) chamber or line
Venous lumen of the catheter obstructed or malpositioned
Blood flow set too high for the catheter
What to check
Check clamps, kinks and patient position
Inspect the deaeration chamber for clot
Flush the venous lumen
Lower blood flow; persistent alarms: assess the catheter
Return pressure too low
Less back-pressure than expected, which can mean blood is not reaching the patient.
A return-line disconnection can cause rapid blood loss. Treat as urgent.
Likely causes
Return line disconnected or leaking
Blood flow lower than set
Pressure line not connected to the pod
What to check
Stop the blood pump and check the return connection
Look for blood at the connection or on the dressing
Pre-filter pressure rising / pressure drop rising
Resistance inside the filter is rising, usually from clotting.
High filtration fraction concentrating blood in the fibers
Clot in the return chamber
What to check
Check anticoagulation: on citrate, post-filter ionized calcium and citrate dose; on heparin, the aPTT or anti-Xa
Review filtration fraction: raise blood flow or move replacement pre-filter
Minimize stops; prepare to change the set before it clots fully
Transmembrane pressure (TMP) high
The membrane is resisting ultrafiltration: fouling or clotting, or too much convection for the blood flow.
Likely causes
Membrane fouling by protein (filter age)
Fiber clotting
Convective flows high relative to blood flow (high filtration fraction)
What to check
Check filtration fraction on the Dose page
Raise blood flow or move replacement pre-filter
Check anticoagulation
TMP that keeps rising usually means the filter is near the end of its life
From the machine readings
mmHg
mmHg
mmHg
mmHg
TMP—mmHg
Filter pressure drop—mmHg
TMP = (P2 + P4) ÷ 2 − P3 · Pressure drop = P2 − P4. Trends matter more than single values: a steady rise in pressure drop points to filter clotting; a rising TMP with a stable pressure drop points to membrane fouling or too much convection.
Typical ranges are approximate and vary with the machine, catheter, blood flow and settings. Follow your device manual and unit protocol.
DISCLAIMER: The authors make no claim of the accuracy of the information contained herein. The calculations suggested above are inherently prone to errors and bias, and should not be a substitute for clinical judgement. The authors involved in the preparation of the site and its content should not be liable for any special, consequential, or exemplary damages resulting in whole or part from any user's use of or reliance upon this material. PLEASE READ THE DISCLAIMER PRIOR TO USING THE SITE. BY ACCESSING AND USING THIS SITE, YOU AGREE TO BE BOUND BY THE TERMS AND CONDITIONS OF THE DISCLAIMER.